PC-2200 User’s manual Advanced Electronic/Air Column PC-2200 BOWERS METROLOGY page 1 /32 version 2.
PC-2200 BOWERS METROLOGY page 2 /32 version 2.
PC-2200 CONTENTS 1. INTRODUCTION 4 1.1 1.2 1.3 1.4 1.5 GENERAL INFORMATIONS TECHNICAL SPECIFICATIONS COLUMN DESCRIPTION KEYPAD CONFIGURATION STATUS PANEL 4 5 6 7 8 2. INSTALLATION 9 2.1 2.2 MECANICAL INSTALLATION ELECTRICAL INSTALLATION 9 10 3. HOW TO USE IT 11 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.
PC-2200 CHAPTER 1. INTRODUCTION 1.1 GENERAL INFORMATIONS - The column PC-2200 is an instrument system that measures and displays “ID, OD, Length related measurements such as misalignment, and others” in automotives, machines, electrical and electronic parts, and determines whether the measured results are acceptable or not.
PC-2200 1.2 TECHNICAL SPECIFICATIONS General : Storage temperature Operational temperature Power supply Max.
PC-2200 1.3 COLUMN DESCRIPTION Rear View Front View ON-OFF Switch Fuse 2A AC power line input Bar scale 3-color LED-bar(103ea) Green, orange, red Auxiliary power line (for other column) Identification and serial numbers RS 232 C Interface External Control Interface OUTPUT Interface (class) Display (LED) (8 chr.
PC-2200 1.4 KEYPAD CONFIGURATION Program Measuring mode Access to the programming mode Programming mode No function Move upward Measuring mode Select M1, M2, M3, M4 Programming mode Used to move from a specific menu to other menu, or to increase the values when setting the values of variables. Escape Measuring mode Initialization of dynamic mode Programming mode Escape to previous menu.
PC-2200 1.5 STATUS PANEL Metric display (mm) Inch display Range ±1000µm / ±1 mm Range Range ±500µm / ±0.5 mm Range Range ±100µm / ±0.1 mm Range Range Range ±50µm / ±0.050 mm Range ± 0.001 inch Range ± 0.0005 inch ± 0.05 inch ± 0.01 inch ± 0.005 inch ±10µm / ±0.01 mm Range Range ± 0.0001 inch ±5µm / ±0.005 mm Measure number BOWERS METROLOGY page 8 /32 version 2.
PC-2200 Chapter 2. Installation 2.1 MECHANICAL INSTALLATION Fix the column on the aluminum base and lock it with the 2 screws (M6 x16) Connect the power supply cable (A) . Connect the probe cable to the analog board (B) Switch (ON) I. A B BOWERS METROLOGY page 9 /32 version 2.
PC-2200 2.4 ELECTRICAL INSTALLATION Switching the Voltage from 230V(110V) to 110V(230V). Remove the cover by unscrewing the 8 screws. Move the connector A on 110V or 230V. A Power supply (PCB) 3PMale Transformer (Default) 220V 0V 110V to noise filter module 2P-female BOWERS METROLOGY page 10 /32 version 2.
PC-2200 Chapter 3. HOW TO USE IT 3.3 SWITCHING ON THE COLUMN To start the first time, we advise to make a general RESET . Press Enter + ON simultaneously, ▓ B O W E R S ▓ Do not release your pressure on program will start. Enter R ES E T . . . - 1 . 0 4 0 0 before the message RESET appears on the display, then a test The default parameters are : CHANNEL : Source : Ext. - CH-A Ratio :1.000 Direc. + MEASURE : M1 – Bar :Bottom B.fact. 1.00 – Range : ± 0.05” Sens : 1.000 Resol.:0.
PC-2200 3.2 PROGRAMMING A SIMPLE MEASURE Example N°1 Measure a component with a HBT probe connected to channel A. Component 20.000mm ± 0.005 USING a HBT probe, we do not use the calibration, this mode will be explained in a subsequent chapter. Master 20.0000mm Step n°1 is to place the probe into a correct position to be able to get the maximum range and the minimum deviation. Place the master 20.0000 mm on the stand, place the probe into position. Press simultaneously Enter S Enter 1 .
PC-2200 3.3 PROGRAMMING A PART (EXAMPLE 1) Press Prog (2 sec.) P / W 0 0 0 0 Enter M 1 Enter S E L E CT B AR R ANG E Enter ± 1 . 0 ± 0 . 5 ± 0 . 0 1 ± 0 . 0 0 5 ± 0 . 0 5 S EN S . R ESO L . Enter Esc 0 . 0 0 0 2 Enter R ESO L . S PE C . Enter M 1 S - M 0 . 8 0 0 0 + TO L ER S M Enter + + TO L ER S Esc 0 . 0 0 4 0 M Enter 0 2 0 . 0 0 0 0 Enter 0 . 8 0 0 0 - AP P RO Enter + S A V E ? S Enter Enter Esc N OM I . Enter Esc R ANG E MEA S URE 0 0 0 .
PC-2200 3.4 PROGRAMMING A DIAMETER Measurement of a diameter with a HBT probe connected to channel A. Component 15.000 mm ± 0.002 Min master 14.9988mm Max. master. 15.0026mm Step n°1 is to place the probe into a location to be able to get the maximum range and the minimum deviation. Place the measuring head of 15.0000 mm in the master, Press simultaneously Enter S Enter 1 . 0 5 0 0 C H - A A D J U S T Move the probe down to the electro-mechanical zero. zero, lock the probe . Press Esc 0 .
PC-2200 3.5 PROGRAMMING A PART (EXAMPLE 2) Press Prog (2 sec.) P / W 0 0 0 0 C HA N NE L Enter S OUR CE Enter D I R E C + Enter R A T I O Enter D I R E C Esc Enter M 1 Enter S E L E CT B AR B . F A CT R ANG E Enter ± 1 . 0 ± 0 . 5 ± ± ± 0 . 0 1 ± 0 . 0 0 5 C H - A - MEA S URE 0 . 1 Enter C H - A 0 . 0 5 C HAN NE L Esc R ANG E S EN S . R ESO L . Enter 0 . 0 0 1 0 . 0 0 0 5 0 . 0 0 0 2 0 . 0 0 0 1 Enter R ESO L . M 1 Esc S PEC . S - M 0 .
PC-2200 The Calibration Example N°2 The parameters are now stored. The automatic calibration can be started. Min. master 14.9988mm Cal C A L . . M1 Enter Max.master 15.0026mm C A L . place the measuring head in the lower master. place the measuring head in the upper master. C A L . Enter Enter 0 . 9 8 1 The sensitivity is calculated automatically. If the message M I N then the calculated value. Cal C A L . .
PC-2200 3.6 THE CLASSIFICATION MENU CLASS The class menu is used separately. Example : Pin dia. 10mm - 0 +0.100mm to be classified by steps of 0.005mm , in 20 classes, starting from 10.0000 mm. Prog (2 sec) P / W Enter S PE C . C L A S S MAS T ER O F F 0 1 0 S T A R T S Enter S T A R T 0 0 0 . 0 0 5 0 ON M Enter OK MEA S URE C HA N NE L Enter S E L E CT 0 2 0 M 0 0 0 . 0 0 0 0 S S T E P Enter S T E P Esc C Enter Enter M S E L E CT Enter N UMB ER Enter N UMB ER 0 1 0 .
PC-2200 3.2 THE RS232 INTERFACE The RS 232 Interface allows to send the data to a printer, as well to a PC, also allows to modify the stored parameters in the column with the help of the communicator software. How to connect a PC-2200 to a serial printer. The most important to connect a column to a printer or a computer is the adjustment of parameters. The first values to be known are the serial printer parameters. e.g. : 4800 bauds, 7 bits, parity even. adjustment How to setup the column ? Prog (2 sec.
PC-2200 3.1 LAYOUT OF IN/OUTPUT 6 7 RS-232 C 8 9 Pin# 1 2 3 4 5 6 7 8 9 Functions * (not connected) Rxd Txd Vcc Gnd * (not connected) Rts Cts Gnd 1 2 3 4 5 6 7 EXT.CON 8 9 Pin# 1 2 3 4 5 6 7 8 9 Functions RS 232 sortie (0.5 sec.) Current value display Dynamic/Static M1->M2->M3->M4 Hold measure * (not connected) * (not connected) * (not connected) Gnd 1 2 3 4 5 6 7 OUTPUT 8 9 Pin# Functions 1 2 3 4 5 1 2 3 4 5 6 7 DC OUT 8 9 1 2 3 4 5 6 7 CH-BUS 8 9 1 2 3 4 5 6 7 8 9 O.
PC-2200 Chapter 4. Functions Definition of keypad functions per menu 4.1 THE MEASURING MODE FUNCTIONS Functions Functions Access to programming mode Prog “M” Change the measure number (M1-M2) only if M2,M3,M4 (ON) M M1-M2-M3-M4 Esc “S” send the displayed value to RS232 S No function in measuring mode No function in measuring mode Enter Preset Cal Preset Value M1 00.0000 Calibration Module M1, M2, M3,M4 CAL% Simultaneously when switching (ON) RESET Access to preset mode Value = Norm.
PC-2200 Prog Menu MEASURE Sub-menu M1 Functions ON Functions BAR B. FACT RANGE SENS. RESOL. FUNC MODE M2, M3, M4 ON/OFF ………. Functions BOTTOM,CENTER,-10->0 1.00 ± 0.05, ±0.01, ±0.005, ± 0.001, ± 0.0005, ±0.0001 1.000 0.005,0.001,0.0005,0.0001,0.00005, 0.00002, 0.00001” A,B,C,D,A+B,A-B,C+D,(A+B)-C, (A+B+C)/3,(A+B+C+D)/4, (A+B)-(C+D) STATIC,MAXIMUM,MINIMUM, (M+m)/2,(M-m)/2,M-m.
PC-2200 Prog Menu CLASS Sub-menu Select Functions ON/OFF Functions NUMBER START STEP Functions 000 +0.0000” 0.0000 Functions Functions 4800,9600,19200 BAUD RATE 7,8 Bits NONE, ODD,EVEN CRLF / CR ON / OFF BRIEF / FULL PRINT /HOST 000 CLASS : Switch ON or OFF the class menu. NUMBER : Store the number of classes to be used. START : Store the starting nominal value STEP : Store the step value of the class.
PC-2200 4.3 CH-BUS APPLICATIONS The CH-BUS channel transfers a data signal to another column. Application. Component 0.002 15.000mm ± 0.007 1 probe We need to see 2 values simultaneously. 1. Diameter 15.0000mm 2. max T.I.R 0.002mm Dia. Max-min Min.master. 14.9988mm Max.master 15.0026mm 4.4 CH-BUS MECHANICAL INSTALLATION Column n°2 Column n°1 Power supply cable Cable PC-2302 The receptive column should have the analog board removed.
PC-2200 4.5 PROGRAMMING THE CH-BUS The column n°1 should be programmed like example n°2 on page n°15 Programming column n°2 P / W Prog 0 0 0 0 C H - B US E X T . C HAN NE L Enter Enter S OUR CE Enter S OUR CE Enter C HA N NE L Esc M 1 Enter S E L E CT B AR B . F A CT R ANG E Enter ± 1 . 0 ± 0 . 5 ± ± ± 0 . 0 1 ± 0 . 0 0 5 MEA S URE Enter 0 . 1 0 . 0 5 R ANG E S EN S . R ESO L . Enter 0 . 0 0 1 0 . 0 0 0 5 0 . 0 0 0 2 0 . 0 0 0 1 Enter R ESO L . F UNC .
PC-2200 Chapter 5. PNEUMATIC INSTALLATION 1. Fix the column and the converter on the aluminum base. 2. Fix the hose of the air probe to nipple of converter. Hose nipple for the air probe 3. Connect the cable PC-2332 between the converter and the analog board. 4. Connect the air supply hose to the hose nipple on the converter. Air supply hose nipple Connection cable PC-2332. Not less than 3 bar of air should be supplied. BOWERS METROLOGY page 25 /32 version 2.
PC-2200 5.1 THE A/E CONVERTER < Front View> (View) (Diagram) The programming procedure is similar to an electronic probe. BOWERS METROLOGY page 26 /32 version 2.
PC-2200 5.2 HOW TO SETUP FOR A/E CONVERTER A. BALANCE B. PRESSURE E. SIGNAL C. GAGE D. AIR IN It is necessary to adjust the A/E CONVERTER in order to use this Column optimally. 1) Pressure Adjustment(STANDARD : 2 ± 0.1bar) - Connect PRESSURE GAUGE to “C”. Connect AIR(Min. 3 bar) to “D”. Adjust pressure by turning “B” with screw driver. ♦ Default pressure is 2 bar(Factory setting). 2) Balance adjustment - Connect the A/E converter and the Column with cable(PC-2332). Connect the air gauge to “C”.
PC-2200 Chapter 6. GENERAL INITIALIZATION (1) Function: Initialize all stored data to the default values. CAUTION : All your stored data will be erased. Press Enter + ON 6.1 TABLE OF DEFAULT VALUES Menu CHANNEL Sub-menu SOURCE Selection EXT. CH-BUS CH-A CH-B CH-C CH-D MEASURE M1 SELECT Functions Choice Default value EXT. RATIO DIREC. RATIO DIREC. RATIO DIREC. RATIO 0 - 5.000 +/0 - 5.000 +/0 - 5.000 +/0 - 5.000 1.000 + 1.000 + 1.000 + 1.000 DIREC. ON BAR +/- + ON BOTTOM B.
PC-2200 6.2 APPLICATION EXAMPLES 6.3 DIAMETER/THICKNESS External diameter : 10.000 mm Lower tolerance : - 20µm Upper tolerance: +20µm Lower warning limit: - 15µm Upper warning limit: +15µm Lower master : 9.980 mm Upper master : 10.020 mm Unit : mm A B Menu CHANNEL Sub-menu SOURCE Selection EXT. CH-BUS CH-A CH-B CH-C CH-D MEASURE M1 SELECT Function RATIO DIREC. RATIO DIREC. 0 - 5.000 +/0 - 5.000 +/- 1.000 + 1.000 + RATIO DIREC. RATIO DIREC.
PC-2200 6.4 BLANK SETTING SHEET Nominal size : Lower tolerance: Upper tolerance: Lower warning limit: Upper warning limit: Lower master : Upper master: Unit : mm Menu CHANNEL Sub-menu SOURCE Selection EXT. CH-BUS CH-A CH-B CH-C CH-D MEASURE M1 SELECT Function RATIO DIREC. RATIO DIREC. 0 - 5.000 +/0 - 5.000 +/- RATIO DIREC. RATIO DIREC. ON BAR 0 - 5.000 +/0 - 5.000 +/- B.FACT RANGE SENS. RESOL. FUNC MODE SPEC.
PC-2200 Chapter 7. General 7.1 TROUBLE SHOOTING Default The digital display & LED display are off. The digital display is off. The digital display & LED display are not moving. The function M2, M3, M4 do not work The calibration is refused “OVER MINIMUM” BOWERS METROLOGY Actions Check the power(AC 110) Check the fuse(2A) Check the power switch(ON) position Reset the column.(“Enter” key + Power ON) Check the probe position Check the connection of probe & column.
PC-2200 7.2 MAINTENANCE Please do not forget that this measuring instrument must be treated with care. Take notice that this “User’s manual” is in relation with the software version V.2.00. Please do not hesitate to contact us if you need further assistance. BOWERS METROLOGY page 32 /32 version 2.
Menu Map for PC2200(V2.xx) BN20914 YES(ENTER) Prog P/W > 2 sec CHANNEL MEASURE SPEC EXT. M1 SELECT CH-BUS PRESET VALUE CH-A M1 ⓐ M2 ⓐ M3 ⓐ M4 ⓐ ON M1 CLASS RS232C RATIO= 0 ~ 5.000 DIREC.= + BAR RATIO= DIREC.= CAL..M(X) MOUDLE CH-EA (xxxx) RATIO= B.FACT 0.1 ~ 5.00 RANGE ⓑ SENS. ±0.010~5.000 RESOL ⓒ FUNC A 0 ~ 5.000 M1 CAL ̄ M2 xxxx M3 xxxx M4 NONE + M2 VALUE 000% 0 ~ 5.000 + C - D " CH-ED - MASTER ⓓ SELECT OFF SPEED ETC.