Product Info
Electromagnetic Compatibility
Vercise™ Deep Brain Stimulation Systems Information for Prescribers
92366224-01 21 of 29
Table 5: Manufacturer’s Declaration
Proximity Fields
The Boston Scientic DBS System is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The users of the Boston Scientic DBS System can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Boston Scientic DBS System as recommended below,
according to the maximum output power of the communications equipment.
Proximity
Test
IEC 60601‑1‑2
Test Level
Compliance
Level
Electromagnetic
Environment Guide
IEC
61000-4-3
385 MHz: 27 V/m
@ 18 Hz pulse modulation
27 V/m
Recommended
separation distance
d = 30 cm
450 MHz: 28 V/m
@ FM modulation
28 V/m
710 MHz, 745 MHz, 780 MHz: 9 V/m
@ 217 Hz pulse modulation
9 V/m
1720 MHz, 1845 MHz, 1970 MHz: 28 V/m
@ 217 Hz pulse modulation
28 V/m
IEC
61000-4-3
1720 MHz, 1845 MHz, 1970 MHz: 28 V/m
@ 217 Hz pulse modulation
28 V/m
Recommended
separation distance
d = 30 cm
2450 MHz: 28 V/m
@ 217 Hz pulse modulation
28 V/m
5240 MHz, 5500 MHz, 5785 MHz: 9 V/m
@ 217 Hz pulse modulation
9 V/m
Note: For the frequency bands in this table, use the specied recommended separation distance. The
recommended minimum separation distance of 30 cm between portable and mobile RF communications
equipment (transmitters) and the Boston Scientic DBS System apply to all other frequencies within the
specied ranges.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reection from structures, objects and people.