Product Info

Electromagnetic Compatibility
Vercise™ Deep Brain Stimulation Systems Information for Prescribers
92366224-01 21 of 29
Table 5: Manufacturers Declaration
Proximity Fields
The Boston Scientic DBS System is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The users of the Boston Scientic DBS System can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Boston Scientic DBS System as recommended below,
according to the maximum output power of the communications equipment.
Proximity
Test
IEC 60601‑1‑2
Test Level
Compliance
Level
Electromagnetic
Environment Guide
IEC
61000-4-3
385 MHz: 27 V/m
@ 18 Hz pulse modulation
27 V/m
Recommended
separation distance
d = 30 cm
450 MHz: 28 V/m
@ FM modulation
28 V/m
710 MHz, 745 MHz, 780 MHz: 9 V/m
@ 217 Hz pulse modulation
9 V/m
1720 MHz, 1845 MHz, 1970 MHz: 28 V/m
@ 217 Hz pulse modulation
28 V/m
IEC
61000-4-3
1720 MHz, 1845 MHz, 1970 MHz: 28 V/m
@ 217 Hz pulse modulation
28 V/m
Recommended
separation distance
d = 30 cm
2450 MHz: 28 V/m
@ 217 Hz pulse modulation
28 V/m
5240 MHz, 5500 MHz, 5785 MHz: 9 V/m
@ 217 Hz pulse modulation
9 V/m
Note: For the frequency bands in this table, use the specied recommended separation distance. The
recommended minimum separation distance of 30 cm between portable and mobile RF communications
equipment (transmitters) and the Boston Scientic DBS System apply to all other frequencies within the
specied ranges.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reection from structures, objects and people.