92366224-01 20 of 29
Table 4: Immunity Testing
RFID Readers
The external electrical components of the Boston Scientic DBS System have been tested for immunity to
interference from RFID readers per the following specications.
RFID Spec Per AIM 7351731 Frequency Test Level (RMS)
ISO 14223 134.2 kHz 65 A/m
ISO/IEC 14443‑3 (Type A) 13.56 MHz 7.5 A/m
ISO/IEC 14443‑4 (Type B) 13.56 MHz 7.5 A/m
ISOAEC 15693 (ISO 18000‑3 Mode 1) 13.56 MHz 5 A/m
ISO 18000‑3 Mode 3 13.56 MHz 12 A/m
ISO/IEC 18000‑7 433 MHz 3 V/m
ISO/IEC 18000‑63 Type C 860‑960 MHz 54 V/m
ISO/IEC 18000‑4 Mode 1 2.45 GHz 54 V/m