Product Info
Electromagnetic Compatibility
Vercise™ Deep Brain Stimulation Systems Information for Prescribers
92366224-01 19 of 29
Table 3: Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
The Boston Scientic DBS System is intended for use in the electromagnetic environment specied below.
The customer or the user of the Boston Scientic DBS System should assure that it is used in such an
environment.
Immunity Test
IEC 60601‑1‑2
Test Level
Compliance Level
Electromagnetic Environment
‑ Guidance
Conducted RF
IEC 61000‑4‑6
(ETS only)
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM and
amateur radio bands
between 150 kHz
and 80 MHz
Professional healthcare facility
environment and home healthcare
environment.
Radiated RF
IEC 61000‑4‑3
10 V/m
80 MHz to 2,7 GHz
10 V/m
80 MHz to 2,7 GHz
Professional healthcare facility
environment and home healthcare
environment.
Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey
a
, should
be less than the compliance level in
each frequency range. Interference
may occur in the vicinity of equipment
marked with the symbol shown below:
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reection from structures, objects and people.
a Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an
electromagnetic site survey should be considered. If the measured eld strength in the location in which
the Boston Scientic DBS System is used exceeds the applicable RF compliance level above, the Boston
Scientic DBS System should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re‑orienting or relocating the Boston Scientic DBS System.