Product Info

Vercise™ Deep Brain Stimulation Systems Information for Prescribers
Vercise™ Deep Brain Stimulation Systems Information for Prescribers
92366224-01 10 of 29
Stimulator Orientation: Orient the Stimulator parallel to the skin surface. Suboptimal
placement of the Stimulator may result in a revision surgery. Patients should avoid touching
the Stimulator site or incisions. If patients notice a change in the appearance of the skin at the
Stimulator location, such as the skin becoming thin over time, they should contact their healthcare
provider.
In order to ensure effective device communications, including device programming, and proper
charging, perform the following:
For rechargeable IPGs, orient the Stimulator parallel to the skin surface and at a depth
less than 2 cm and greater than 0.5 cm below the skin.
For non-rechargeable Vercise Genus IPGs, orient the Stimulator parallel to the skin
surface and at a depth less than 2.5 cm below the skin to ensure effective device
communication.
For the non‑rechargeable Vercise PC IPG (DB‑1140‑S), orient the Stimulator parallel to
the skin surface. There is no depth restriction for the Vercise PC IPG.
The etched writing “This Side Up” must be facing out of the pocket toward the patient’s
skin.
Suboptimal placement of the Stimulator may result in the inability to communicate with the device
or inability to recharge and may require a revision surgery.
Patients should be instructed not to change the orientation of or turn over the Stimulator. If the
Stimulator ips over in the body, then it cannot communicate or be charged. If stimulation cannot
be turned on after charging, the Stimulator may have changed orientation or rotated; patients
should contact their healthcare provider to arrange an evaluation of the system.