User's Manual

Speci cations and Technical Data
Implantable Pulse Generator
90655621-03 Rev A 3 of 11
Speci cations and Technical Data
Parameter Range Default
Areas (Channels) 4
Amplitude 0 – 25.5 mA 0 mA
Rate 2 – 1200 pps
a
40 pps
Width 20 – 1000 sec
b
210 sec
Cycle 0.1 – 90 min, OFF OFF
Ramp ON 1 – 10 secs 3 secs
Contacts 1 – 32, case: +100% to -100%, OFF 1 – 32, case: OFF
a. Only one Area is available if the rate is 130 pps.
b. Amplitude × Width 12.7 C.
Materials
Case Titanium
Header Epoxy
Strain Relief Silicone
Size/Volume 55.0 mm x 46.0 mm x 10.8 mm / 21.2 cm
3
(including header)
Radiopaque Identi cation Tag
The IPG contains a radiopaque identi cation tag “BSC IPG”. The identi cation tag is visible
using standard x-ray procedures.
Registration Information
In accordance with international practice and regulatory legislation in some countries, a
registration form is packed with each Boston Scienti c neurostimulator. The purpose of this
form is to maintain traceability of all products and to secure warranty rights. It also allows the
institution involved in the evaluation or replacement of a speci c device to gain quick access to
pertinent data from the manufacturer.
Fill out the registration form included in the package contents. Return one copy to Boston
Scienti c, keep one copy for patient records, provide one copy to the patient, and one copy to
the physician.