User's Manual

Instructions for Use
Implantable Pulse Generator
90655621-03 Rev A 7 of 11
CAUTION: Do not tunnel splitter.
Note: If using the 2x8 Splitter and performing a permanent trial, the splitter tails may be
tunneled to the exit site.
Note: The following Codman Disposable Catheter Passers may be used in place of the
Boston Scienti c tunneling tool:
REF 82-1515 (36 cm); REF 82-1516 (55 cm); REF 82-1517 (65 cm)
Note: When using a Codman Disposable Catheter Passer, tunnel from the midline incision to
the IPG pocket using the standard technique.
Connecting the Lead, Extension, Splitter, or Connector to the IPG
IPG ports are labeled as follows:
For convenience, connect leads or splitter tails to the IPG ports corresponding to their locations,
superior versus inferior or left versus right lead placements. For example:
Superior leads to upper IPG ports A or B. Inferior leads to IPG ports C or D.
For the In nion™16, connect the splitter tail with laser etched bands (contacts 1-8 of the
In nion 16 lead) to the left ports A or C and the unmarked splitter tail (contacts 9-16 of
the In nion 16 lead) to the right ports B or D.
For the Artisan 2x8 surgical lead, connect the left side to the left ports A or C. Connect
the right side (the laser-etched tail), contacts 9-16, to the right ports B or D.