Charging the Implant
31
4. Locate the Charger over the implant. When aligned with the implant, the beeping will stop.
Secure the Charger over the implant either by pressing the adhesive side to the location, or by
attaching the Velcro
®
belt.
Note: If you accidentally locate the patch in the wrong place, or if the Charger moves out of
alignment, the Charger will start beeping again. Use a new adhesive patch or readjust the
Ve lc ro
®
belt to place the Charger back to position.
5. When the Charger emits a distinct double
beep, the implant is charged. Switch off the
Charger, remove the adhesive patch or
Velcro
®
belt, and return the Charger to the
Base Station.
Do not confuse the end of charge (a distinct
double beep) signal with the steady, contin-
uous misalignment signal.