Guidelines for Trial-phase Implantation
25
6. Grasp the tip of the tool with one hand while holding the straw in
place with the other hand. Pull the tunneling tool shaft out
through the straw.
7. Push the lead or extension proximal ends through the straw, then
withdraw the straw.
8. Wipe clean the proximal end of the lead, then insert the proximal
end into the extension connector until it stops and the last contact
disappears.
Note: If there appears to be an obstruction, use the torque wrench to
loosen (counterclockwise) the setscrew and/or gently rotate
the lead to help advance the proximal end.