NurseCall Main Unit Appendix | en 83
TeleAlarm SA® User Manual 953.92 | v1.0 | 2007.07
13.5.4 Other paging systems
Only Alarms/Messages arriving to the Alarm Buffer are transmitted (except "RELAY OFF"). The
transmission is fully alpha-numeric. The Alarms/Messages are repeated every 1 or 3 minutes
until Acknowledgement (see Section 7.4.13 RS232 message setting (toggle), page 55).
The Alarms/Messages are transmitted to the paging system user number attributed to the
Transmitter (default group = 00).
i
NOTICE!
Multitone Access 3000 compact, etc. can be connected if configured correctly
(RPE 670 = NO, ESPA 4.4.4.).
i
NOTICE!
Data rate: 9600 Bauds. Transmission: asynchronous with a 10 bit-structure
(1 startbit, 7 databits with even parity, 1 stopbit) in half-duplex mode.
Pagingsystem-ID = 2, NurseCall-ID=1.
Group User number
00 999
01 998
02 997
03 996
04 995
05 994
06 993
07 992
08 991
09 990
10 989
11 988
12 987
13 986
14 985
15 984
16 983
17 982
18 981
19 980
20 979
21 978
Table 13.9 Attribution table