54 en | Programming NurseCall Main Unit
953.92 | v1.0 | 2007.07 User Manual TeleAlarm SA®
Each type of Transmitter has its own ID code range. Regarding the type, the unit adds an offset
to the ID code:
Procedure
1. Type the code 001999.
2. Confirmation message displayed.
3. Set the CONVERSION CODE OFF or ON with
the keys
and .
X If you select ON, the offset are ignored.
Then, the information sent out for
Transmitters RAC, N46 and S37 which have
the same ID code, is identical.
4. Set the LAST 300 BLOCKED function OFF or
ON with the keys
and .
X If you select ON, the last 300 ID from
previous Transmitters which can reach 6560
(respectively 13121 for RAC-Fire) are not
managed!
7.4.9 Maximum number of Alarm Transmitters (toggle)
This command allows to set the maximum number of Alarm Transmitters
(300=OFF or 500=ON).
The default value is ON.
1. Type the code 001001.
2. Confirmation message displayed.
Transmitter ID code range Offset Data sent
S37, S35 1 to 4095 0 1 to 4095
S36-RAC-Old version 0 to 6560 0 0 to 6560
RAC-Fire 1 to 4095 6561 6562 to 10656
RAC-Fire-Old version 0 to 6560 6561 6561 to 13121
N45, N46 1 to 4095 13122 13123 to 17217
Table 7.3
UNIVERSAL NC? YES !
............
CONVERSION CODE ?
<OFF ON>
LAST 300 BLOCKED ?
<OFF ON>
Enter Code
Parameter ON