User manual
NurseCall Main Unit Programming | en 51
TeleAlarm SA® User Manual 953.92 | v1.0 | 2007.07
7.4 Special settings
After pressing the key three times quickly, you
can enter the following special codes.
7.4.1 Displaying firmware version
1. Type the code 194155.
2. The version of the firmware will be displayed
for a few seconds.
7.4.2 Resetting all parameters (including date and time)
The following procedure is used to reset all programmed parameters of the NurseCall Main
Unit to the original factory settings.
X See also Section 7.2.1 List of original factory settings, page 37.
1. Type the code 194156.
2. A confirmation is required. Press the
key
to confirm the reset or
if you wish to
cancel the reset.
3. After a few seconds, you have to select the
language, set the Localization Mode ON or
OFF and select the Display Mode, as for the
first use.
X See Section 7.2 First use
Enter Code
* TELEALARM NCALL3 *
V1.00 BN111.140.00A
i
NOTICE!
It is mandatory to perform this command if you wish to change the values for the Localization
Mode and the Display Mode.
X See also Section 7.2.3 Localization Mode, page 38.
X See also Section 7.2.4 Display Mode, page 38.
i
NOTICE!
It is mandatory to perform this command before setting the NurseCall Main Unit as Universal
NurseCall or as Standard NurseCall.
X See also Section 7.4.7 Standard NurseCall selection, page 53.
X See also Section 7.4.8 Universal NurseCall selection, page 53.
RESET TOTAL ?
(OK) = YES