50 en | Programming NurseCall Main Unit
953.92 | v1.0 | 2007.07 User Manual TeleAlarm SA®
7.3.7 Local Acknowledgement setting
This parameter setting is used for the setting of the Local Acknowledgement.
1. Select the parameter No. 03.
2. Press
.
3. Set NO if Acknowledgement at the NurseCall
Main Unit shall be disabled or set YES if you
wish to enable this function.
4. Confirm the setting with
.
5. If you have selected YES, you should enter
the access code.
X Press the key
then .
6. Select YES if each Acknowledgement must be
done by entering the code 45 or NO if direct
Acknowledgement with the Green button
shall be enabled.
7. Confirm the setting with
.
7.3.8 Output relay setting
This parameter setting is used for relay programming as "closing" or "switching" contact.
1. Select the parameter No. 04.
2. Press
.
3. Select ON if you wish a "closing" relay or
ON/OFF if you wish a "switching" relay.
4. Select relay activation according to Alarms.
You have 3 possibilities:
– Mode HELP & ASSIST;
– Mode ASSISTANCE;
–Mode FIRE;
5. Confirm the setting with
.
X Parameter Nr. 03
Local Ack. ¡
X Local Ack.
Possible ? YES ¡
ACCESS CODE ··
X Local Ack.
Access Code ? YES ¡
X Parameter
Stored !
X Parameter Nr. 04
Relay Output ¢
X Relay Output
Fonc: 'ON'? ¡
X Relay Output
Mode HELP & ASSIST¡
X Parameter
Stored !