User manual
46 en | Programming NurseCall Main Unit
953.92 | v1.0 | 2007.07 User Manual TeleAlarm SA®
7.3.6 RS-232 output setting
Parameter setting for RS-232 interface programming.
1. Select the parameter No. 02.
2. Press .
X Set the desired value with the
and
keys.
3. Confirm the setting with .
X For detailed paging or DECT phone system programming, see Section Example of
programming, page 47.
i
NOTICE!
Select one of the following:
–None
–Printer
– Alarm Management SW
–PAGING
–DECT
–S10-Du
X Parameter Nr. 02
RS-232 Output ¡
X RS-232 Output
Printer ¡
X Parameter
Stored !
i
NOTICE!
For the values None, Printer and Alarm Management SW, you do not have to define more
parameters. If you connect a paging or a DECT phone system, more parameters are available
i
NOTICE!
Inside the NurseCall Main Unit, the RS-232 interface should be configured with jumpers.
X See Section 6.2.7 Setting the jumpers on RS-232 board, page 32.