4.2.5 RS-232 interface
A 9-pole SUB-D connector at the rear of the housing can be used for connection to a
–printer;
– paging system;
– DECT phone system;
–PC with Alarm Management Software.
Connection to a printer
To protocol all Events, a printer with serial connection (RS232-Interface) and endless paper
should be used. Printers with a parallel port can be used together with an intermediate serial
- parallel converter.
Data rate: 9600 Bauds. Transmission: asynchronous with a 10 bit-structure
(1 startbit, 8 databits without parity, 1 stopbit).
The availability of the printer cannot be tested (switched on/off, paper).
The following printer is recommended:
–SCRIPTOS.
Connection to a paging system
The NurseCall system uses several protocols, standard ESPA 4.4.4. protocol, a specific
Multitone protocol (Access 700-MEP) as well as the DeTeWe protocol.
X See Section 13.5 Paging systems specifications, page 78 for more information about these
protocols.
Connection to a DECT phone system
The NurseCall system can transfer the received Alarms to DECT handsets Multitone CH60 or
CH70.
X See Section 13.6 DECT phone system specifications, page 85 for more information about
this system.
i
NOTICE!
X For the hardware configuration of this interface, see Section 6.2.6 Connecting the RS-232,
page 31.
X For the programming of this interface, see Section 7.3.6 RS-232 output setting, page 46.
X For the wiring of the connector, see Section 13.7.1 LINE socket (unit bottom), page 86.
i
NOTICE!
The paper printout corresponds to the indication at the display of the NurseCall Main Unit.