Operator`s manual

32
LightCycler® 480 Gene Scanning Software
Performing Gene Scanning Analysis
Click Sample Editor in the Module bar, then select the workflow Scanning. After the
workflow is selected, the LightCycler 480 Software automatically reconfigures the
Sample Editor.
Define the properties of the samples.
For detailed information on the Sample Editor see the LightCycler
®
480 Instrument
Operator’s Manual, section “Entering Sample Information”.
The software uses the following parameters for calculation:
Column Name Description Valid Values
Target Name
Name of the (PCR) target
(e.g., the name of the gene
amplified and detected)
Alphanumeric value
( 25 characters);
default value is
“Target xxx-yyy”,
where xxx and yyy
are the excitation and
emission wavelengths
Sample Type
Type of sample
If you want to use the
“In-run Standards”
grouping option, you need
to include a Melting Stan-
dard for any genotype you
want to detect.
Choose from dropdown list:
Unknown
Negative Control
Melting Standard
Genotype
Genotype of melting standard
or positive control samples
This field is active only
when the Sample Type is
“Melting Standard”.
Alphanumeric value
( 25 characters)
Click Analysis in the Module bar. In the Create New Analysis list, select Gene Scanning.
► ► ►