Table 2-2. Peak Power Sensor Characteristics (con't.)
Model
Frequency Power Overload
Rise Time
Maximum SWR
Drift & Noise
Range Measurement Rating
@ 0 dBm
Peak Fast Slow
Impedance
CW
(1)
Peak Power High Low Frequency SWR Peak Power
RF Connector Int. Trigger CW Power Bandwidth Bandwidth CW Power
(GHz) (dBm) (ns) (ns) (GHz)
DUAL DIODE PEAK POWER SENSORS
Sensors below are for use with 4400, 4500, 4400A, 4500A and 4530.
Compatible with 4530 Series internal 50 MHz calibrator.
57318 0.5 to 18 -24 to 20 1W for 1 us
< 15
(2)
< 10 us to 2 1.15 4 uW
50 Ω
(0.05 to 18) -34 to 20 200 mW (35 MHz) (350 kHz) to 16 1.28 0.4 uW
N(M) -10 to 20 to 18 1.34
(3)
57340 0.1 to 40 -24 to 20 1W for 1 us
< 15
(2)
< 10 us to 4 1.25 4 uW
50 Ω
(0.03 to 40) -34 to 20 200 mW (35 MHz) (350 kHz) to 38 1.65 0.4 uW
K(M) -10 to 20 to 40 2.00
(3)
57518 0.1 to 18 -40 to 20 1W for 1 us < 100 < 10 us to 2 1.15 50 nW
50 Ω
(0.05 to 18) -50 to 20 200 mW (6 MHz) (350 kHz) to 6 1.20 5 nW
N(M) -27 to 20 to 16 1.28
(4)
to 18 1.34
57540 0.1 to 40 -40 to 20 1W for 1 us < 100 < 10 us to 4 1.25 50 nW
50 Ω
(0.05 to 40) -50 to 20 200 mW (6 MHz) (350 kHz) to 38 1.65 5 nW
K(M) -27 to 20 to 40 2.00
(5)
NOTES: 1) Models 4400, 4500, 4400A and 4500A only.
2) Models 4531 and 4532: <20ns, (20MHz).
3) Shaping Error (Linearity Uncertainty), all levels 2.3%
4) Shaping Error (Linearity Uncertainty), all levels 4.0%
5) Shaping Error (Linearity Uncertainty), all levels 4.7%
Fre
q
uenc
y
calibration factors
(
NIST traceable
)
and other data are stored within
all the Peak Power Sensors. Linearit
y
calibration is
p
erformed b
y
the built-in
calibrator of the
p
eak
p
ower meter.
MODELS 4400
,
4500
,
4400A and 4500A:
All Peak Power sensors can be used with these models and calibrated with the
internal 1GHz ste
p
calibrator unless otherwise noted.
MODELS 4531 and 4532:
The Peak Power sensors in the lower
g
rou
p
above ma
y
be used with these models
and calibrated with the internal 50 MHz ste
p
calibrator. The sensors on the u
pp
er
g
rou
p
ma
y
be used if the Model 2530 1 GHz Accessor
y
Calibrator is used for
calibration.
A five-foot lon
g
sensor cable is standard. Lon
g
er cables are available at a hi
g
her
cost. Effective bandwidth is reduced with longer cables.