6. Fault Diagnosis
Symptom
Action Required
1. Unit does not operate
a. Switch on
b. Plug in, switch on
c. Replace fuses per 8.2
d. Check that other
electrical appliances on
the same circuit are
working
2. Chamber temperature
does not rise when
expected
a. Check set temperature
b. Have unit checked by
competent person
c. Turn timer off.
3. Temperature continues
to rise when not
expected
a. Check setting
b. Have unit checked by
competent person
4. No rotation, Display
illuminated
a. Turn Carousel switch on
b. Remove obstruction
c. Tighten set screws w/hex
key
Catastrophic Heater Condition: In the event that the chamber temperature
would reach 120°C the systems heater will be turned OFF, and will on longer heat the
chamber. Also the heat indicator lamp on the control panel will remain illuminated and
an Alarm will sound for a short period of time. To reset the unit back to normal operation
cycle the power (OFF/ON) at the units power switch on the rear of the oven.
IF THIS CONDITION PERSISTS HAVE THE UNIT CHECKED BY A
COMPETENT PERSON BEFORE CONTINUING THE USE OF THE
HYBRIDIZATION OVEN!