Illuminated Tissue Flotation Bath Model 145702 145702-2 145951 145951-2 Operating Instructions N2400313 Rev.
TABLE OF CONTENTS 1. 2. 3. 3.1 3.2 4. 4.1 4.2 4.3 4.4 4.5 4.6 4.7 5. 6. 7. 8. 8.1 8.2 9. 10. Page Safety .............................................................................................................................. 2 Product Information ........................................................................................................ 2 Assembly......................................................................................................................... 3 Unpacking .......
1. Safety The following symbols marked on the equipment mean: Caution: Read these operating instructions fully before use and pay particular attention to sections containing this symbol. Attention: Suivre attentivement les instructions avant l’usage et prêtez une attention particulière aux sections comportant ce symbole. Caution: Surfaces can become hot during use. Attention: Les surfaces peuvent devenir brûlantes pendent l’usage. Always observe the following safety precautions: • • • • • • • • • • • • 2.
The Tissue Bath has a water temperature range of ambient + 5°C to approximately 65°C under normal operating conditions. The unit has an Orienter Block which has a fixed temperature of 55°C ± 3°C and a Dryer Block which has a fixed temperature of 40°C ± 3°C. 3. 3.1 Assembly Unpacking Remove packing materials carefully, and retain for future shipment or storage of the unit. Inspect for damage. Complete and return the Warranty Registration Card.
Heating Indicator Alarm Indicator FIGURE 1 4.2 Display Up Adjustment Button Down Adjustment Button Filling the Bath Fill the glass dish with water to a minimum depth of 2cm. Caution: Water should be placed only in the glass dish. Filling the chamber without the glass dish will damage the bath and may pose an electrical hazard.
4.4 Setting High Temperature Alarm 4.4.1 The unit is programmed with an adjustable high temperature alarm. The high temperature alarm value is factory set at 5⁰C above the set temperature value. To change this value, follow the steps below. 4.4.2 Depress the “Alarm” button; the display will now alternate between “Alar” and the current high temperature alarm value. Depress the “Set” button and the display will now flash the current high temperature alarm value.
5. Fault Diagnosis Symptom 1. Unit does not operate 2. Bath temperature does not rise when expected 3. Temperature continues to rise when not expected 4. Lamp not illuminated 6. Possible Cause a. Unit not switched on b. Unit not plugged into power supply c. Fuses blown d. Power supply failure a. Actual temperature is higher than set temperature b. Temperature control sensor is not positioned in the water c. Temperature control circuit fault a. Actual bath temperature is lower than set temperature b.
. Maintenance and Service All Boekel laboratory products are designed to comply with IEC1010-1. No routine maintenance is required. There are no user serviceable parts in this product. 8.1 Cleaning Disengage power cord prior to cleaning. The glass dish has molded handles to allow removal for easy cleaning. Paraffin build-up should be avoided around the top area of the Tissue Bath.