Boekel Incubator Models 132000, 132000-2 Operating Instructions N2400222 Rev.
CONTENTS Page 1. Safety ................................................................................................................. 2 2. Product Information ............................................................................................ 3 3. Assembly............................................................................................................ 3 3.1 Unpacking ..................................................................................................... 3 3.
1. Safety Always observe the following safety precautions: • • • • • • • • • • • • Use only as specified by the operating instructions or the intrinsic protection may be impaired. After transport or storage in humid conditions, dry out the unit before connecting it to the supply voltage. During drying out the intrinsic protection may be impaired. Connect only to a power supply with a voltage corresponding to that on the serial number label.
2. Product Information The Boekel series of Incubators are designed for laboratories and clinics that require contemporary styling, accuracy and economy in a basic incubator package. Boekel’s 132000 Incubator provides excellent incubator performance at the most economical cost. This Incubator uses a dial controlled thermostat for temperature control and provides a top enter ferrule entry point for thermometer insertion. The 132000 incubator has a chamber volume of 0.8 cu. ft. (0.02 cu.
132000 SHELF INSTALLATION DIAGRAM STEP #1 STEP #2 INNER WALL SHELF BRACKET INNER WALL 90° STEP #3 INNER WALL INNER WALL SHELF 4. Operation 4.1 Controls and Indicator Lamps (see Figure below) The power switch controls power to the unit. The heater lamp indicates when the heater is operating. The temperature adjust knob is used to control the chamber temperature.
4.2 Setting the Temperature Use the temperature adjust knob to set the desired temperature. The knob is rotated clockwise to increase the chamber temperature and counter clockwise to decrease the chamber temperature. 4.3 Operating the Incubator To turn the incubator on, switch the power switch to the on (I) position. Set the desired chamber temperature. The heater lamp will illuminate and remain lit until the chamber reaches the set temperature.
6. Fault Diagnosis Symptom 1. Unit does not operate 2. Chamber temperature does not rise when expected 3. Temperature continues to rise when not expected Possible Cause a. Unit not switched on b. Unit not plugged into power supply c. Power supply failure Action Required a. Switch on b. Plug in, switch on a. Actual temperature is higher than set temperature b. Temperature control circuit fault a. Check set temperature a. Actual temperature is lower than set temperature b.
. Maintenance and Service All Boekel laboratory products are designed to comply with IEC1010-1. No routine maintenance is required. There are no user serviceable parts in this product. 8.1 Cleaning Disengage power cord prior to cleaning. If a spill occurs, use appropriate clean up procedures as required for radiation or biohazard control. The outer casing may be cleaned with water and a damp cloth. Do not submerge or immerse the Incubator in water.