Chapter 4 Common Modulation Output RIGOL
User‟s Guide for DG5000
4-13
To Set Modulating Amplitude
Press ModAmp to set the modulating amplitude.
Input the desired amplitude value using numeric keyboard or knob.
The available amplitude (High Z) ranges from 0 V to 10 V, and the default value is
2 V.
To Set Modulating Polarity
Press Polarity to select the Positive or Negative of the modulating waveform to
control the output amplitude.
When internal modulation is selected, if the polarity is positive, logic low of the
modulating waveform amplitude corresponds to output of the greater of the two
amplitudes (modulating amplitude and carrier amplitude), while logic high
corresponds to output of the smaller amplitude; if the polarity is negative, the
situation is the opposite.
When external modulation is selected, if the polarity is “positive”, logic low of the input
corresponds to output of the greater of the two amplitudes (modulating amplitude
and carrier amplitude), while logic high corresponds to output of the smaller
amplitude; if the polarity is “negative”, the situation is the opposite.