Leading/Trailing Edge Time
The Leading (rising) edge time is defined as the duration of the pulse amplitude rising
from 10% to 90% threshold, while the Trailing (falling) edge time is defined as the
duration of the pulse amplitude moving down from 90% to 10% threshold.
Press Pulse Leading (or Trailing ) to highlight the softkey, and then use the
numeric keyboard to input “50” and select the unit “ns” in the pop-up menu.
Edge Time range: 2.5 ns to 1 ms.
Selectable Leading/Trailing Edge Time units: sec, msec, μsec, nsec.
You can also use the knob to modify this parameter.
Delay
Delay is defined as the delayed time between main output and sync output.
Delay
Snyc Output
Main Output
Press Pulse Delay to highlight the softkey, and then use the numeric keyboard to
input “8” and select the unit “μs” in the pop-up menu.
Delay Time range: 0 s to pulse period
Selectable Delay Time units: sec, msec, μsec, nsec.
You can also use the knob to modify this parameter.
Recover Delay
Press Pulse, use to open the 2/2 menu page and select Restore, the
generator will restart to output pulse waveform after specified delay time.