User Manual

RIGOL Chapter 10 Advanced Operation
User‟s Guide for DG5000
10-34
To Set the Code Pattern
Press Pattern to select the code pattern of the output data: ALL_0, ALL_1, 01, 8PRBS
(Pseudo Random Binary Sequence), 16PRBS, 32PRBS or User (user-defined code
pattern).
When Useris selected, users can also set the starting Offset of the data output. The
range of the offset is from 0 to 262143 Bytes. Offset is relative to the address,
Offset=1 represents data stored in the first address (1 byte data) will not be output.
Note, the setting of the offset does not change the mask channel data under PO
protocol. The first mask is always the data stored in the address 00000000 and
corresponding to the mask channel set. For example, when the mask channel is set to
D3 and the data in the address 00000000 is 0x08, the first mask is 1.