User Manual

RIGOL Chapter 10 Advanced Operation
User‟s Guide for DG5000
10-8
Channel Coupling
The dual-channel model of DG5000 supports coupling output of frequency or phase.
When coupling is enabled, set CH1 or CH2 as the “base” channel and set the
frequency or phase deviation of the two channels. Then when the frequency or phase
of the base channel is changed, the corresponding parameter of the other channel will
change automatically and always keeps the specified frequency or phase deviation.
Press Utility at the front panel, use to open the 2/2 menu page and press
Coupling to enter the Channel coupling interface.
Press Coupling to turn On or Off the coupling function. The default is Off.
Press Type to select FreqDev or PhaseDev, and use the numeric keyboard to enter
the desired value. Here, PhaseDev is set as 10 °.
Set Coupling as Off, and press Base to select CH1 or CH2 as the coupling base
source. Here, CH2 is selected. Then phase relationship between CH1 and CH2 is:
P
CH1
=P
CH2
+ 10 °
When the coupling base source is under the Freq/Phase menu mode, the screen will
display a * mark. The figure below shows that the instrument is in phase coupling
mode currently. If the phase of CH2 is changed to 0 °, the phase of CH1 will be
automatically adjusted to 10 °.
Figure 10-2 Channel Coupling
Coupling Mark