User Manual

Chapter 10 Advanced Operation RIGOL
User‟s Guide for DG5000
10-5
Sync Polarity
Set the sync signal on the [Sync] connector as normal or invert.
Press Utility CH1 Set Polarity to select Positive or Negative. Notice that
the sync signal related to the waveform is not invert even when the waveform is
invert.
Output Polarity
Set the signal on the [Output] connector as normal or invert. Waveform inverts
relatively to the offset voltage.
Press Utility CH1 Set Output to select Normal or Invert. The default is
Normal. The first half period of the cycle waveform is positive in Normal mode and
negative in Invert mode.
0 V Offset
0 V
Normal Invert Normal Invert
After the waveform is inverted, please pay attention to the following cases
None offset voltage is changed.
The waveform viewed in Graphic mode is not invert.
The sync signal related to the waveform is not invert.