User Manual
RIGOL Chapter 9 Store and Recall
User‟s Guide for DG5000
9-10
Seamlessly Interconnection
DG5000 could interconnect with RIGOL oscilloscope which supports USB-TMC
seamlessly and rebuilt and output the waveform data collected from the oscilloscope
lossless.
1. Make sure the measured signal is displayed stably on the screen of oscilloscope
DS1.
2. Connect the USB Host interface of DG5000 to the USB Device interface of DS1.
Now, DS1 will enter “Remote” mode automatically.
3. Press Store/Recall key at the front panel of DG5000 to enter Store/Recall
interface. The model identifier of DS1 will appear in the “Directory Display Area”
(refer to Figure 9-1).
4. Use the knob of DG5000 to move the focus cursor onto the model identifier of DS1.
All channels and their on/off status of DS1 will be displayed in the “File/Folder
Display Area”.
5. Set the Browser of DG5000 as “File”, and move the cursor onto the channel to be
read, and then press Read. DG5000 will automatically switch to arbitrary
waveform mode (Arb) and load the waveform collected by DS1 to the internal
volatile memory of DG5000. Meanwhile, in order to protect the measured circuit,
DG5000 will automatically turn off its output. When the user confirms that the
current output will not damage the measured circuit and turns on the channel
switch of DG5000, the generator will output the arbitrary waveform collected by
DS1. Connect the output of DG5000 to the input of DS2 and configure DS2 to view
the waveform collected by oscilloscope DS1.
Tips: The waveform data collected by the oscilloscope is stored circularly, thus the
data read by DG5000 is also finite and exists discontinuousness in phase. We
recommend that you display interested signal fragment in the oscilloscope screen
stably, and then use DG5000 to read the data and then output through burst (Burst)
mode.
DG5000
DS1
DS2
USB Device USB Host OUTPUT INPUT
USB Cable BNC Cable