RIGOL Chapter 6 Frequency Sweep Output
User‟s Guide for DG5000
6-8
The generator will renewedly sweep and output with the specified Start Frequency
once the Mark Frequency is changed.
Start Hold
Start Hold is the duration that the output signal outputs with the Start Frequency
after the sweep starts. After the Start Hold, the generator will output with varying
frequency in the current sweep type.
When Sweep is enabled, use to open the 2/2 menu page, then press Start
Hold and use the numeric keyboard or the knob to change the Start Hold. The default
value is 0 s and the available range is from 0 s to 300 s.
The generator will renewedly sweep and output with the specified Start Frequency
once the Start Hold is changed.
End Hold
End Hold is the duration that the output signal outputs with the End Frequency after
the generator has swept from the Start Frequency to the End Frequency.
When Sweep is enabled, use to open the 2/2 menu page, then press End
Hold and use the numeric keyboard or the knob to change the End Hold. The default
value is 0 s and the available range is from 0 s to 300 s.
The generator will renewedly sweep and output with the specified start frequency
once the End Hold is changed.