Log Sweep
The output frequency of the instrument varies in a logarithmic fashion, that is, the
output frequency changes in the way of octave per second or decade per second.
The change is controlled by Start Frequency, End Frequency and Sweep Time.
When Log Sweep is enabled, users could set the following parameters:
Start Frequency
start
F
, End Frequency
stop
F
and Sweep Time
sweep
T
The function prototype of Log Sweep is
T
current
PF
. P and T could be expressed by
the above-mentioned parameters:
sweepstartstop
TFF
P
/)/lg(
10
)lg(/)lg( PFtT
start
Where, t is the time from the start of the sweep and its range is from 0 to
sweep
T
.
current
F
is the instantaneous frequency of the current output.
When Sweep is enabled, press SwpType to select Log. In the Graphic Mode, an
exponential function curve, which indicates the output frequency changes in a
logarithmic mode, will appear on the screen (shown in the figure below).