Instruction Manual
Chapter 3 Arbitrary Waveform Output RIGOL
DG4000 Series User’s Guide
3-13
Example: Edit Points
The following example introduces how to use “Edit Points” to create an arbitrary
waveform fulfilling the following conditions.
Operation Procedures:
1. Press Arb and use to open the 2/2 menu page; then press Create New
Æ Cycle Period, use the numeric keyboard to input “12” and select the unit
“μs” from the pop-up menu.
2. Press High VLimit, use the numeric keyboard to input “4” and select the unit
“V” from the pop-up menu.
3. Press Low VLimit, use the numeric keyboard to input “-2” and select the unit
“V” from the pop-up menu.
4. Press Init #Points , use the numeric keyboard to input “4” and press OK.
Note: at this point, a -2 V level line appears.
5. Press Interp to select “Linear”.
6. Press Edit Points to enter the “Edit Points” interface.
Parameter
Value
Cycle Period
12μs
High VLimit
4V
Low VLImit
-2V
Interp Linear
Point 1 0s, 0V
Point 2 4μs, 4V
Point 3 8μs, 0V
Point 4 10μs, -2V
BNC Scientific, a division of Berkeley Nucleonics Corporation
2955 Kerner Blvd, San Rafael CA 94901 USA –www.bncscientific.com
Chat Online or email info@bncscientific.com– Toll Free: 866-276-6188, Fax 707-262-6522