Note: enable high-frequency reject when low-frequency signal with lower
than 1kHz frequency is measured to filter out high-frequency noise
interference; disable high-frequency reject when high-frequency signal with
greater than 1kHz frequency is measured.
7. Gate Time
Press Gate Time to set the gate time of the measurement system and the
default is 1.310ms.
GateT1 1.310ms
GateT2 10.48ms
GateT3 166.7ms
GateT4 1.342s
GateT5 10.73s
GateT6 >10ns
8. Measure
Press Measure to select the type of parameter to be measured by the counter.
The counter can measure the following parameters: frequency (Freq),
period (Peri), duty cycle (Duty), positive pulse width (+Width) and
negative pulse width (“–Width) and the default is Freq.
9. Statist
Press Statist to enable or disable statistic function. For detail information,
please refer to Statistic.
10. Auto
Press this key to set the counter parameters automatically.
After the counter parameters are set, the generator measures the external input
signal using the preset parameters.