Manual

48
Model 835-M Programmer’s
Manual
V1.1 2011
[:SYSTem:COMMunicate] Subsystem
Command Parameters
Unit
(defa
ult)
:SYSTem:COMMunicate:LAN:CONFig DHCP|MANual|AUTO
DHC
P
:SYSTem:COMMunicate:LAN:DEFaults V0.4.1
:SYSTem:COMMunicate:LAN:DHCP:TIMeout N/A 30 sec
:SYSTem:COMMunicate:LAN:GATeway
<ipstring>
“0.0.0.0”
:SYSTem:COMMunicate:LAN:IP
<ipstring>
:SYSTem:COMMunicate:LAN:RESTart
:SYSTem:COMMunicate:
LAN:S
U
B
N
et
<ipstring>
“255.255.255
.0”
:SYSTem:COMMunicate:SOCKet:ECHO
ON|OFF|1|0 OFF
:LAN:CONFig
:SYSTem:COMMunicate:LAN:CONFig
DHCP|MANual|AUTO
:SYSTem:COMMunicate:LAN:CONFig?
This command sets
the
s
ignal
generator’s
internet
protocol (IP)
address.
MANual The
user assigns
an
IP address
to the
signal
generator.
DHCP
The network
assigns
an
IP address
to the
signal
generator.
If
DHCP
fails,
manual configuration will be used.
AUTO The network
assigns
an
IP address
to the
signal
generator
with
a
fallback
to Auto-
IP
if
DHCP
fails. If both
DHCP
and Auto-
IP
fail, manual
c
onfiguration
will be used.
:L
AN:DEF
aults
:SYSTem:COMMunicate:LAN:DEFaults
This command restores
the
instrument’s LAN settings
to their factory default values.
:LAN:DESCription
(not implemented)
:SYSTem:COMMunicate:LAN:DESCription
<string>
:SYSTem:COMMunicate:LAN:DESCription?
This command
defines the
instrument’s
web
description.
The query returns the current saved
setting.
:LAN:DHCP:TIMeout
(not implemented)
:SYSTem:COMMunicate:LAN:DHCP:TIMeout
{30}|60|90|120sec
:SYSTem:COMMunicate:LAN:DHCP:TIMeout?
This command enables
the
user
to
change
the maximum length of time that the instrument
will
spend
trying to
acquire
an
IP address using DHCP.
If the
LAN
Config Type
is
set to Auto,
then
the Auto-
IP protocol
will be
used as
a fall-
back
when time- out
does occur.
The
DHCP
ti
meout
value
is
stored in the
same
non- volatile ram
as
the other
LAN configurations.
The query
retur
n
s
the current
setting, not the
saved
setting.
Default 30
S
e
conds
:LAN:DOMain
(not implemented)
:SYSTem:COMMunicate:LAN:DOMain
<string>
:SYSTem:COMMunicate:
LAN:
DOMain
?