Manual

27
Model 835-M Programmer’s
Manual
V1.1 2011
[SOURce]:POWer:ALC:HOLD
ON
|OFF
|1|0
[
S
OU
Rce]
:POWer:AL
C
?
(Devices
with option
PE
only)
This command sets
the automatic level control into hold mode. The amplitude level control loop
is
open.
*RST
The preset value
is
0
:ATTenuation
[SOURce]:POWer:ATTenuation
<value> [SOURce]:POWer:ATTenuation?
(Devices
with option
PE
only)
This command sets
the power range extension attenuator.
This command
will
also
turn
off
automatic
attenuation setting. Refer to
[SOURce]:POWer:ATTenuation:AUTO
OFF
for details. Attenuator input
RF
power
shall
be
-30…+10 dBm,
e.g.
using
a
[SOURce]:P
O
Wer:AT
Tenu
ation
value of 50
dB,
the output
RF
power range
is
-80…-40 dBm.
*RST
The preset value
is
0
:ATTenuation:AUTO
[SOURce]:POWer:ATTenuation:AUTO ON|OFF|1|0
[SOURce]:POWer:ATTenuation:AUTO?
(Devices
with option
PE
only)
This command
turns the power range extension on or off.
Turning
it off allows fast
power
sweeps
for
devices
featuring an extended output power range.
*RST
The preset value
is
0
:ATTenuation:LIST?
[SOURce]:POWer:ATTenu
a
tion:AUTO?
(Devices
with option
PE
only)
This
query
returns
a
comma-separated
list of available attenuation
settings. These can
be set
using
the
[SOURce]:POWer:ATTenuation
command.
Available
settings
are:
0, 35, 70
dB
for option
PE
2A
0,
10,
20, …, 70
dB
for option PE3
0,
10,
20, …,
110 dB
for option PE4
0, 35, 70,
105 dB
for option PE5