Manual
11
Model 835-M Programmer’s
Manual
V1.1 2011
The
Service Request Enable (SRE) command sets
the value of the
Service Request
Enable
Register.
The
variable <data>
is
the
decimal sum
of the bits that will be enabled. Bit 6 (value 64)
is
ignored and
cannot be set by this command.
Range
0–255
The setting enabled by this
command is
not affected by
signal
generator preset or
*RST.
However,
cycling
the
signal
generator power will reset it to zero.
*SRE
?
The
Service Request Enable (SRE)
query returns the value of the
Service Request
Enable
Register.
Range
0–63 & 128-191
*STB
?
The
Read Status
Byte
(STB)
query returns the value of the
status
byte
including
the master
summary status (MSS)
bit.
Range
0–255
*TRG
The Trigger
(TRG) command triggers
the
device
if
LAN is
the
selected
trigger
source,
otherwise,
*TRG
is
ignored.
*TST?
The
Self-Test (TST)
query initiates the internal self- test and returns one of the following results:
0
This shows
that all
tests
passed.
1
This shows
that one or more
tests
f
ailed.
*WAI
The Wait- to-
Continue
(WAI)
command causes
the
signal
generator to wait until all pending
commands
are
completed,
before
executing
any other commands.