Owner manual
Table Of Contents
- Parts List
- Specifications
- Basic Functions
- How You Will Use the Model 725
- Installing timerPRO Software
- Setting Up the Model 725
- Field Upgradeable Software
- Getting Help
- Back Panel Inputs and Outputs
- Front Panel Controls
- Main Screen
- General Settings
- Channel Properties Screen
- Quiescent States
- Logic
- Timing Modes
- “Apply” versus “Program All Channels”
- Fixed Output Mode
- Passive Mode (Output = Input)
- Clocked Pulse Stream Mode
- Delayed Pulse After Trigger Mode
- Important Delay Timing Considerations
- Validated Trigger Input Mode (Noise Suppression)
- Timer Mode
- Dynamically-Delayed Pulse Mode
- Toggled Output Mode
- Dynamic Delay Compensation Mode
- Storing and Recalling Experiment Files
- Front Panel Controls and Indicators
- Selecting a Channel from the Front Panel
- Performing a Front Panel Function
- Store/Recall Settings Profiles

31
Delayed Pulse After Trigger Mode
Upon receiving an input signal, a delayed pulse after trigger waits a specified delay time, then
sends a pulse to the channel output. This mode has several property tabs.
Delay
On the Delay tab you’ll specify:
Delay after trigger The delay between a valid trigger and an output signal
Duration of pulse The duration of the output signal
Until Reset The output will remain high after a valid trigger, until the channel is reset
Inverted Output Check the box to set the output high until a valid trigger.
The range of settings you can use with the delayed trigger depend upon the current timing mode: