Owner manual
Table Of Contents
- Parts List
- Specifications
- Basic Functions
- How You Will Use the Model 725
- Installing timerPRO Software
- Setting Up the Model 725
- Field Upgradeable Software
- Getting Help
- Back Panel Inputs and Outputs
- Front Panel Controls
- Main Screen
- General Settings
- Channel Properties Screen
- Quiescent States
- Logic
- Timing Modes
- “Apply” versus “Program All Channels”
- Fixed Output Mode
- Passive Mode (Output = Input)
- Clocked Pulse Stream Mode
- Delayed Pulse After Trigger Mode
- Important Delay Timing Considerations
- Validated Trigger Input Mode (Noise Suppression)
- Timer Mode
- Dynamically-Delayed Pulse Mode
- Toggled Output Mode
- Dynamic Delay Compensation Mode
- Storing and Recalling Experiment Files
- Front Panel Controls and Indicators
- Selecting a Channel from the Front Panel
- Performing a Front Panel Function
- Store/Recall Settings Profiles

18
Logic Assignments
The logic processor assigns the combination of external inputs (in1–in8) and channel outputs
(outA–outG) that will trigger each channel. Logic assignments always appear in the following
format:
[the channel to be triggered] = [the conditions that will trigger it];
External inputs and channel outputs feed into the logic processors; therefore, they appear only on
the right side of assignments. Logic processor outputs feed the timing channel inputs inA–inH;
these terms appear only on the left side of assignments.
Here is an example:
inB = in2;
This assignment means, “Channel B will be TRUE when Input 2 goes TRUE (reaches ~ 5 V).”
The assignment:
inB = not in2;
means “Channel B will trigger when Input 2 goes FALSE (drops below 0.2 V).”
You can employ compound arguments in logical assignments, such as:
inB = not outC and in2;
This means, “Channel B will trigger when the output of Channel C is FALSE and Input 2 is
TRUE.”
This example is more complex still:
inB = (in1 or in2) and (outC or outD);
Here, Channel B requires a TRUE signal from Inputs 1 or 2, combined with a TRUE signal from
Channels C or D, in order to trigger.
NOTE: Model 725 inputs are TRUE by default if nothing is connected (i.e., the input is
normally weakly pulled high). Grounding an input switches its state to a logical FALSE.