User guide

67
You can set up the RTD configuration either through the front panel
operation or through the remote interface operation.
Front Panel Operation
If you are using RTD, press CONFIG and then TEMP. Use and to
locate SENSOR submenu. Press ENTER to go to the submenu. Use
and to locate your sensor type Press ENTER to select your desired
sensor type.
Choosing USER takes you to a menu where you can specify factors used
in the calculation equation to obtain the temperature. Use and to
move through the digits and and to change the numbers to a
desired value. Press ENTER to set the value.
Choosing SPRTD takes you to a menu where you can specify the seven
coefficients that are used to determine the temperature. Use and
to move through the digits and and to change the numbers to a
desired value. Press ENTER to set the value.
Remote Interface Operation
Use the following commands to set up the RTD configuration.
SENSe:TEMPerature:RTD:TYPE{PT100|D100|F100|PT385|PT3916|USER|SPRTD|NTCT}
SENSe:UNIT {Cel|Far|K}
SENSe:UNIT?
SENSe:TEMPerature:RTD:RZERo {<value>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:ALPHa {<value>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:BETA {<value>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:DELTa {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:RZERo {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:A4 {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:B4 {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:AX {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:BX {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:CX {<value>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:DX {<value>|MINimum|MAXimum}
Thermocouple