Manual
Model 1120 - 1pS Time Interval Counter
NUT 012
Ed. 6
Date : 18/01/2011
BERKELEY NUCLEONICS Page -17 / 36 -
4. CONNECTING TO THE INSTRUMENT VIA ETHERNET
4.1. Command structure
*IDN?
Syntax: *IDN?
Form: Query
Description: Queries instrument identification. Response gives instrument
model, serial number and firmware version.
Parameter: -
Example: GFTy/MIPSI,BNC 1120,SN101/000000,V1.0
Instrument model: BNC 1120
Serial number: 101,
Firmware version: 1.1
ARM
Syntax: ARM START|STOP
ARM?
Form: Set & Query
Description: Define the arming mode or ask for current arming mode.
Command
Parameter:
START arm on start
STOP arm on stop
Query response :ARM START arming on start
:ARM STOP arming on stop
Example: ARM START
ARM? => :ARM START