Datasheet

TSYS01-FAMILY
Digital Temperature Sensor
SENSOR SOLUTIONS /// TSYS01-FAMILY
06/2018
Page 5
COMMANDS
The commands are the same for SPI and I
2
C interface.
There are four commands:
Reset
Read PROM (calibration parameters)
Start ADC Temperature conversion
Read ADC Temperature result
Command
Hex Value
Reset
0x1E
Start ADC Temperature Conversion
0x48
Read ADC Temperature Result
0x00
PROM Read Address 0
0xA0
PROM Read Address 1 (Coefficient k
4
)
0xA2
PROM Read Address 2 (Coefficient k
3
)
0xA4
PROM Read Address 3 (Coefficient k
2
)
0xA6
PROM Read Address 4 (Coefficient k
1
)
0xA8
PROM Read Address 5 (Coefficient k
0
)
0xAA
PROM Read Address 6 (SN
23...8
)
0xAC
PROM Read Address 7 (SN
7...0
and Checksum)
0xAE
RESET SEQUENCE
The Reset sequence has to be sent once after power-on. It also can be used to reset the device ROM
from an unknown condition.
SPI
I
2
C