Specifications

5-4
Maintenance TOS9220/9221
Symptom2: The POWER ON LED is lit, but the scanner does not
function.
Symptom3: Pressing the START switch with the contact check
function activated results in an immediate FAIL judg-
ment.
Possible cause Remedy
The SCAN terminal of the tester proper is
connected to the SCANNER OUT terminal of
the scanner.
The interface cable is not connected correctly. Connect it
correctly by referring to "3.1 Connecting the Scanner
to the TOS9200/9201 Tester".
A channel not configured on the TOS9200/9201
tester is being used.
Channels connected to test points must be set on the
tester proper.
For information on the procedure for setting channels,
see “Channel settings for the high-voltage scanner ” in
Chapter 3 “Basic Operations” of the Operation Manual
for the TOS9200/9201 Tester.
Possible cause Remedy
A high-voltage lead for parallel connection is
disconnected.
Connect the high-voltage leads properly by referring to
"3.1 Connecting the Scanner to the TOS9200/9201
Tester".
The OUTPUT A and B terminals are not
connected to the corresponding test point, or a
wire is disconnected in any of the connections
(TOS9220 only).
Install or reconnect the leads properly and securely by
referring to "3.3 Connecting the TOS9221 Scanner to
the DUT".