Specifications
5-2
Maintenance
TOS9220/9221
5.3
Replacing the Fuse
•
T
o pre
v
ent electr
ic shoc
k, bef
ore chec
king or replacing the fuse
, be sure to
tur
n off the PO
WER s
witch of the tester proper and unplug the scanner’
s
A
C po
w
er cord.
•
Mak
e sure that the fuse used conf
or
ms to the instr
ument specifi
cations
,
including shape
, r
ating, and char
acter
istics
.
Using a fuse with diff
erent r
at
-
ing or shor
t-circuiting, the fuse holder will damage the instr
ument.
1.
T
ur
n off the PO
WER s
witch of the tester proper and unplug the scan
-
ner’
s A
C po
w
er cord from the electr
ical outlet.
2.
Disconnect the A
C po
w
er cord from the A
C LINE connector on the rear
panel.
3.
Use a slotted-head scre
w dr
iv
er to remo
v
e the fuse holder
, as sho
wn in
Fig.
5-1
.
Fig.5-1
Remo
ving the Fuse Holder
T
ab
le5-1
Fuse Rating
•
The pre-arcing time-current characteristic of fuses are named dif
ferently in the UL
and IEC standards. Use fuses conforming to both or either of the standards.
WARNING
Spare fuse
Fuse
200-240V
47-63Hz
40
100-120V
NOTE