Specifications
TOS9220/9221 Part Names and Functions 4-3
[7] OUTPUT A (TOS9221 only)
The A-output terminal of each channel is an output terminal on the high-voltage
side (HIGH VOLTAGE).
The A-output terminal of a channel set to HIGH on the TOS9200/9201 tester is
connected to the high-voltage side (HIGH VOLTAGE potential) of the test volt-
age. The A-output terminal of a channel not set to HIGH is in the open state (not
connected to another point).
[8] OUTPUT B (TOS9221 only)
The B-output terminal of each channel is an output terminal on the low-voltage
side (LOW).
The B-output terminal of a channel set to LOW on the TOS9200/9201 tester is
connected to the low-voltage side (LOW potential) of the test voltage. The B-out-
put terminal of a channel not set to LOW is in the open state (not connected to
another point).
When the tester proper has activated the contact check function, voltage is tempo-
rarily applied to the B-output terminals of channels set to both the high-voltage
and low-voltage sides, in order to check the contact condition. Since in this wir-
ing configuration high voltage is also applied to the four terminals provided as the
low-voltage side output terminals, they should be handled in the same way as the
A-output terminals.
[9] IN HIGH VOLTAGE
This terminal connects to the OUT HIGH VOLTAGE terminal of the tester proper
or to another TOS9220/9221 scanner connected in parallel, in order to receive the
test voltage from the TOS9200/9201 tester. During a test, the test voltage is sup-
plied continuously.
[10] IN LOW
This terminal connects to the LOW terminal of the TOS9200/9201 tester or to the
OUT LOW terminal of another TOS9220/9221 scanner connected in parallel.
[11] OUT HIGH VOLTAGE
This terminal supplies the test voltage supplied from the tester proper to the IN
HIGH VOLTAGE terminal of another TOS9220/9221 scanner connected in paral-
lel when multiple scanners are used in parallel connection. During a test, the test
voltage is supplied at all times.
• This terminal is not used to connect to the DUT.
[12] OUT LOW
This terminal connects to the IN LOW terminal of another TOS9220/9221 scan-
ner when multiple scanners are used in parallel configurations.
NOTE