Service manual

7 义݅9
PRELIMINARY INFORMATION ----- SUBJECT TO CHANGE
DŽ
Chroma2329 Fluke5418 CA-210
3. ADC Calibration
a. Signal and generator
PC: Chroma2329 Pattern42, Timing 14(1024x768@60Hz 5 MOSAIC)
YPBPR: Chroma2329 Pattern103, Timing 79(100% Color Bar, 720p)
5 MOSAIC 100% Color Bar
b. Steps of Calibration:
1. Access to the Factory menu->ADC
2. Select RGB source and see the 5 MOSAIC appears on the screen
3. Select AUTO ADCˈPress the right button on your remote to calibrate the
ADC automatically.
4. Select YPBPR source and Repeat step 3.
When the OK appears the calibration has been finished successfully.



VGA calibration HDMI 1 WB adjustment
4. White Balance adjustments (Manual)