User manual
Blackrock Microsystems CereStim R96 - User Manual
Rev. 1.00 ® Page vi
LB-0432
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
an electromagnetic site survey
a
, should be less than the
compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
** Modulation scheme can be 2 Hz or 1 kHz.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the CereStim R96 is used exceeds the applicable RF compliance level above, the
CereStim R96 should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the CereStim R96.
b
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3 V/m