User manual
Blackrock Microsystems CereStim R96 - User Manual
Rev. 1.00 ® Page 13
LB-0432
Update: New max values are stored by clicking the ‘Update’ button. This button is only available to
users with administrative rights.
Read: All users are able to read the existing Max Value settings by clicking the ‘Read’ button regardless
of administrative rights.
Change Password Panel (3): After unlocking Administrator Rights, users can change an existing password
to a new password by entering the old password into the ‘Enter Old Password’ box, entering the new
password in the ‘Enter New Password’ box and retyping the new password in the ‘Confirm New Password’
box. New passwords must not be more than 32 characters.
Change Password: Click the ‘Change Password’ button to apply changes to the password.
3.4 Saving and Exporting Information
Stim Manager is able to save any changes the user makes in the software environment, including:
activated and deactivated current modules, test electrodes results, pulse configurations, stimulation
scripts, and reports.
Stim Manager is then able to export any of that information as a text file. The default roots are:
Environment files: [application folder]/Documents/
Test electrodes result: [application folder]/Measurements/
System advanced configurations: [application folder]/SystemFiles/
3.5 Application Programming Interface (API)
For customers who want to control the CereStim R96 via proprietary software, a complete API is
made available. Please refer to the CereStim R96 API User Manual included in the CereStim R96
packaging for further information.
4 Device Care and Storage
Handle the CereStim R96 with care. Do not drop or strike the device. In the event that the device is
dropped, the CereStim R96 must be returned to Blackrock Microsystems for inspection and
recalibration before the device can be used again (See section 7).
Clean the outside of the CereStim R96 using a cloth dampened with water or alcohol when
necessary. Do not submerge the device in water or cleaning solutions.
Keep the device away from liquids. The CereStim R96 is not fluid resistant and can become
damaged by fluid ingress.
Store the CereStim R96 at -20˚C to 50˚C, 10 to 100% relative humidity (non-condensing).
5 Troubleshooting
Stim Manager does not recognize the CereStim R96. I am unable to send commands from Stim Manager
to the CereStim R96.
1. Make sure that your PC recognizes the CereStim R96 as an HID interface. Right Click on ‘My
Computer’ in the Start-up Menu and select Manage. Select Device Manager, expand the
Human Interface Devices List and confirm that your USB device is recognized. If the USB is
not recognized, the CereStim R96 may require a shorter USB cable.
2. Check to make sure that all connections are secure.