3.3 Stim Manager Administration Tab
Unlock Administrator Rights Panel (1): Unlocking the Administrator Rights allows users to restrict the
Output Vmax, Amplitude, Phase Width, Interphase Delay, Interpulse Delay and Charge per Phase to a set of
maximum values. Upon initial start-up, the administrator password is set to ‘Stim100.’ After unlocking the
administrator rights, the user can change this password in the ‘Change Password’ panel.
Unlock: After entering the appropriate password, click the ‘Unlock’ button to gain Administrative
Rights.
Lock: After any settings are changed, click the ‘Lock’ button to store the changes and prevent other
users from making adjustments to the settings.
Max Values Panel (2): Only users with Administrative Rights are able to make adjustments to the
maximum allowable values for the Output Vmax, Amplitude, Phase Width, Interphase Delay, Interpulse
Delay and Charge per Phase. When making changes to the Max Values, administrative users must specify
all max values in order for Stim Manager to accept the update. Upon initial start-up, the default Max
Values are as follows:
Max Output Voltage: 6.5V
Max Amplitude: 100µA
Max Width: 300µs
Max Interphase: 300µs
Max Interpulse: 300µs
Max Charge per Phase: 30,000pC
Figure 3-12: Stim Manager Administration Tab