Rev. 1.00 ® Page 5
LB-0432
WARNING: Always make sure that the CereStim R96 is turned off before
connecting/disconnecting any cables
2.4.1.4 Connect the external power supply to the CereStim R96 and plug it into a wall
outlet.
2.4.1.5 Connect the USB port on the Front Panel of the CereStim R96 to a USB port on the
computer with the USB A-B cable (included in packaging).
2.4.1.6 Connect the appropriate end of the CereStim R96 Interconnection Cables to the
CereStim R96 cable connectors labelled A-C on the back panel of the CereStim R96.
These connections are keyed and lock the cables in place when connected
appropriately.
2.4.1.7 Connect the opposite end of the cables to the ‘STIM IN’ Connectors on the front
panel of the Stim Switch Control Module. Secure each connector in place using two
thumbscrews (provided in Stim Switch packaging). Please refer to the Stim Switch
User Manual for further instructions regarding the use of this system.
2.4.1.8 Open the Stim Manager Software on your PC and turn on the CereStim R96. The
status bar in the lower right-hand corner of the Stim Manager window should
change from green to orange, indicating that the CereStim R96 is connected. Refer
to Section 3 below for further instructions regarding the use of the Stim Manager
Software.
WARNING: The CereStim R96 must be turned off when not in use.
2.4.2 CereStim R96 Connection to Electrodes
For customer supplied connections, connection to the CereStim R96 is as follows:
2.4.2.1 Unpack all components of the CereStim R96 packaging.
2.4.2.2 Turn on your IEC 60950 PC and load the Stim Manager Software (included in
packaging).
2.4.2.3 Confirm that the CereStim R96 is turned off.
WARNING: Always make sure that the CereStim R96 is turned off before
connecting/disconnecting any cables
2.4.2.4 Connect the external power supply to the CereStim R96 and plug it into a wall
outlet
2.4.2.5 Connect the USB port on the Front Panel of the CereStim R96 to a USB port on the
computer with the USB A-B cable (included in packaging).
2.4.2.6 Connect one end of the customer supplied cables to the CereStim R96 Cable
Connectors on the Back Panel of the CereStim R96. Connect the other end of the
cables directly to the connectors on the patient electrodes. Refer to Figure 2-4 for
a pinout of the locking connector on the CereStim R96.