User manual

Stim Switch User Manual
Rev. 2.00 ® Page iii
LB-0357
Use only the Front End Amplifier and Neural Signal Processor (NeuroPort™ System) supplied by
Blackrock Microsystems. Substitution of systems not supplied by Blackrock may affect system
performance and patient safety.
The Stim Switch and Amplifier/NSP should be disconnected from the patient electrodes during
defibrillation. The Stim Switch and its subcomponents, along with the Amplifier and NSP are not
defibrillator proof.
Do not use the Stim Switch in the presence of flammable anesthetic agents.
Only connect Stim Switch components to properly tested, grounded and dedicated AC outlets to
reduce the risk of electrical shock. Do not use an adapter for ungrounded wall outlets.
Do not connect the Stim Switch to an outlet controlled by a wall switch.
Avoid strong static discharges from sources like televisions or computer monitors because it can
damage the electrical parts of the system.
Keep the Stim Switch away from liquids. Contact with water, shower spray, or wet surfaces can lead
to the patient receiving an electrical shock.
Each component of the system must be plugged into a separate outlet (i.e. Stim Switch, Stim100,
subcomponents of the data acquisition system and all other accessory equipment). Using a power-
strip may cause leakage currents to exceed acceptable limits.
Do not unplug the power supply to the Stim Switch while the system is in use.
Do not leave the Stim Switch connected to the patient electrodes when the system is not in use.
When connecting the Stim Switch to the Stim100 or patient electrodes, use caution to minimize the
likelihood of Electrostatic Discharge (ESD). Do not connect/disconnect any cables to/from the Stim
Switch while the system is in use.
Ensure that the Stim Switch is securely positioned on a flat surface during use.
Use caution when placing power cords, cables, and other connectors to minimize the likelihood of
tripping or accidentally pulling on cables. Damaged cables may results in failed ground
connections.
Connecting equipment to and from the Stim Switch may result in a summation of leakage currents
that can lead to the patient receiving an electric shock.
Risk of Electrical Shock Hazard: Accessory equipment connected to the SIP/SOP port must be
certified according to the respective IEC standards, i.e., IEC 60950 for data processing equipment
or IEC 60601-1 for electromedical equipment in order to reduce the risk of electrical shock. All
combinations of equipment must be in compliance with IEC 60601-1-1 systems requirements.
Anyone connecting additional equipment to the SIP/SOP port configures a medical system, and
therefore is responsible that the system complies with the requirements of the system standard IEC
60601-1-1.
IEC 60950 approved Information Technology Equipment must be placed outside the “patient
environment.” The patient environment is defined as an area 1.5m (4.92 feet) from the patient.
The following options are recommended if there is a need to remediate system leakage current: