Specifications
Model Name
Cerebus® Neural Signal Processor
Power Requirements
110 (at 60 Hz) or 240 (at 50 Hz) VAC, 8.0 Amps maximum load
Serviceable Fuses
5 x 20mm, 250V, 1.6A, Slow Blow
Compliance Standards
IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26, CSA listed
Type of Protection
Class I (Amplifier Power Supply)
Degree of Protection
Type CF Applied Part (Neural Signal Amplifier and Patient Cable Assembly)
Mode of Operation
Continuous
Water Ingress Protection
Ordinary Equipment, not fluid resistant, IP20
Operating Environment
10˚C to 40˚C, 5 to 95% R.H. (non-condensing)
Storage Environment
-20˚C to 50˚C, 5 to 100% R.H. (non-condensing)