User`s manual

Cerebus® User’s Manual
Revision 13.00 / LB-0028 Page 47
File Storage Options
Click on View and then Options to set file storage options.
Figure 41 - File Storage Options
o Enable Sync Pulse: Will send a sync on Digital Output 1. The sync signal is a unique waveform pattern sent every 14
seconds and repeats itself every hour. Each 14 seconds the pattern changes. This can be used to synchronize recordings
with external equipment.
o Invert Sync Pulse: It will invert the sync pulse. The pulse will start high and will go low.
o Remove Noise: Enabling this option will remove noise-classified waveforms (unit 255) from the recorded file.
o Sync with NeuroMotive Recording: It will enable syncing with the NeuroMotive Video Recording and Tracking
System, if present.
o Split File Interval: The recorded file will be split every so often. The user can enter either a duration or a file size that
the split should occur at. If file size is entered click on “Calculate Time” to automatically populate the “Estimated
Time” field.