Cerebus® User’s Manual
Revision 13.00 / LB-0028 Page 31
Digital Out
This window will allow the user to set the properties of the digital outputs transmitted to other equipment connected
to the NSP.
Figure 26 - Digital Output Properties
o Label: A user-defined name for the specific channel.
o Properties: The user has the option to Monitor Electrodes, which will send a TTL pulse from that specific Digital
Output every time a spike is detected on a specific channel or unit of a channel. As a second option, the user can select
Timed to output a user specified frequency range with a specified duty cycle and offset. The low and high samplings
show the number of clock cycles the square wave is set to high or low. The offset unit is clock cycles and can offset the
trigger of one channel from the other output channels. Each clock cycle is 1/30,000 second long.
o Track Most Recently Selected: When this box is checked the channel source will change based on whichever channel is
selected from Spike Panel, Single Neural Channel, or Raster Plot.
o Source: Select a source channel to monitor.
o Monitor Units: The user can select which units to monitor from the source channel through the digital output ports. The
user can monitor unclassified units and/or classified units defined online.